Researcher Collab

Investigating the Dosimetric Leaf Gap Correction Factor of Mobius3D Dose Calculation for Volumetric-modulated Arc Radiotherapy Plans

The DLG correction factor strongly influences the accuracy of Mobius3D-calculated doses. Applying the optimal DLG correction factor can increase dose agreement and gamma passing rate between calculation and delivered doses of VMAT plans, which emphasizes the importance of optimizing this factor during the commissioning process.

DOI: https://doi.org/10.4103/jmp.jmp_11_24

Publish Year: 2024